Transcutaneous surgical procedures performed with a high intensity focused ultrasound (HIFU) therapy system can be monitored in real-time with an ultrasound imaging system if the HIFU is gated appropriately. Without synchronization, gated or continuous HIFU saturates the imaging system and interference occludes the image. If a gating signal is synchronized with the imaging cycle from any commercial imager, the location of any interference can be controlled and the HIFU treatment region can he visualized in real-time. Synchronization typically requires that an imaging system be customized for HIFU therapy, which is expensive and time consuming. We have developed a low-cost prototype system that synchronizes a HIFU therapy system with an arbitrary unmodified imaging system by using the HIFU transducer as a focused receiver that can detect scattered acoustic signals transmitted by the imaging probe. The receive signal is processed into a trigger that is used to control the gating and phasing of the HIFU relative to the imaging cycle. The technique is tested using a B-mode imager to monitor the formation of a lesion in a transparent tissue phantom; exposure time is 60 seconds with 40 W (time-averaged) electrical power delivered to the transducer. Performance is evaluated by recording the position of interference on the B-mode images and by comparing the B-mode images with CCD images that provide an optical view of lesion formation.
机译:如果对HIFU进行适当的门控,则可以使用超声成像系统实时监控使用高强度聚焦超声(HIFU)治疗系统执行的经皮手术程序。没有同步,门控或连续HIFU会使成像系统饱和,并且干扰会遮挡图像。如果门控信号与任何商用成像仪的成像周期同步,则可以控制任何干扰的位置,并且可以实时显示HIFU治疗区域。同步通常需要为HIFU治疗定制成像系统,这既昂贵又耗时。我们已经开发了一种低成本的原型系统,通过将HIFU换能器用作聚焦接收器,可以将HIFU治疗系统与任意未修改的成像系统同步,该聚焦接收器可以检测由成像探头传输的散射声信号。接收信号被处理成触发器,用于控制HIFU相对于成像周期的门控和定相。使用B模式成像仪测试该技术,以监测透明组织体模中病变的形成。传递到换能器的功率为40 W(时间平均),曝光时间为60秒。通过记录B模式图像上的干涉位置并将B模式图像与提供病变形成光学视图的CCD图像进行比较,来评估性能。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号