首页> 外文会议> >Estimation of pressure head and flow rate in a continuous-flow artificial heart-in vivo evaluation
【24h】

Estimation of pressure head and flow rate in a continuous-flow artificial heart-in vivo evaluation

机译:连续流人工心脏体内评估中压力头和流速的估计

获取原文

摘要

Instead of direct measurement of pressure head and flow rate of a centrifugal blood pump, the authors previously proposed a method of estimating these measurements. In this study, to ascertain the adequacy of the proposed method, an acute animal experiment was carried out using an adult goat equipped with an artificial heart consisting of two centrifugal pumps. The results indicated that the ARX model identified even in a mock circulation could estimate flow rate in the recipient accurately enough, although estimation accuracy of pressure head was not high.
机译:代替直接测量离心血泵的压头和流量,作者先前提出了一种估计这些测量值的方法。在这项研究中,为了确定所提出的方法的适当性,使用配备有由两个离心泵组成的人造心脏的成年山羊进行了一项急性动物实验。结果表明,尽管压头的估计精度不高,即使在模拟循环中识别出的ARX模型也可以足够准确地估计出接收器中的流速。