首页> 外文会议> >Cantilever-based bio-chemical sensor integrated in a microliquid handling system
【24h】

Cantilever-based bio-chemical sensor integrated in a microliquid handling system

机译:集成在微液体处理系统中的基于悬臂的生化传感器

获取原文

摘要

The cantilevers have integrated piezoresistive readout which, compared to optical readout, enables simple measurements on even non-transparent liquids, such as blood. First, we introduce a simple theory for using piezoresistive cantilevers as surface stress sensors. Then, the sensor fabrication based on conventional microfabrication is described and the sensor characterization is discussed. During the characterization we found a stress sensitivity of (/spl Delta/R/R)=4.6:10/sup -4/ (N/m)/sup -1/ and a minimum detectable surface stress change of 2.6 mN/m. Aqua regia etch of gold on top of the cantilevers has been monitored, and immobilization of single-stranded thiol modified DNA-oligos has been detected by the sensor. Finally, it is demonstrated that it is possible to analyze two samples simultaneously by utilizing the laminar flow in the microliquid handling system.
机译:悬臂具有集成的压阻读数,与光学读数相比,它甚至可以对非透明液体(例如血液)进行简单测量。首先,我们介绍一种使用压阻悬臂作为表面应力传感器的简单理论。然后,描述了基于常规微加工的传感器制造,并讨论了传感器的特性。在表征过程中,我们发现应力敏感度为(/ spl Delta / R / R)= 4.6:10 / sup -4 /(N / m)/ sup -1 /,最小可检测表面应力变化为2.6 mN / m。已经监测了悬臂顶部金的王水蚀刻,并且传感器检测到了单链硫醇修饰的DNA-寡核苷酸的固定化。最后,证明了通过利用微液体处理系统中的层流可以同时分析两个样品。