AC leakage currents cause complete hemodynamic collapse below the ventricular fibrillation threshold

机译:交流泄漏电流导致低于心室纤颤阈值的完全血液动力学崩溃

获取原文

摘要

In 1993, the American standard for leakage current through the heart under a single-fault condition was increased from 10 /spl mu/A to 50 /spl mu/A rms. Both the original 10 /spl mu/A standard and the new 50 /spl mu/A standard were based on estimates of the ability of 60 Hz AC current to induce ventricular fibrillation (VF). This basis was a logical choice, since VF causes systemic blood pressure collapse, which can quickly cause morbidity and death. However, Swerdlow et al. (1999) showed that systemic pressure collapse can result from 60 Hz AC leakage currents (84/spl plusmn/27 /spl mu/A rms) far below those required to induce VF (278/spl plusmn/226 /spl mu/A rms) in 32 closed-chest humans. In this work, we extend the human results to a dog model and show preliminary evidence for the mechanism of the collapse.
机译:1993年,单次故障条件下通过心脏的漏电流的美国标准从10 / spl mu / A rms提高到50 / spl mu / A rms。最初的10 / spl mu / A标准和新的50 / spl mu / A标准都是基于对60 Hz交流电引起心室纤颤(VF)的能力的估计。此基础是一个合理的选择,因为VF会导致全身性血压下降,从而迅速导致发病和死亡。然而,Swerdlow等。 (1999年)表明,由于60 Hz交流泄漏电流(84 / spl plusmn / 27 / spl mu / A rms)可导致系统性压力崩溃,远低于诱发VF所需的电流(278 / spl plusmn / 226 / spl mu / A rms)。 )在32个封闭的胸部中进行。在这项工作中,我们将人类结果扩展到狗模型,并显示了崩溃机理的初步证据。