首页> 外文会议> >An inexpensive phantom for evaluating gated blood pool data acquisition/processing systems at heart rates above 400/min
【24h】

An inexpensive phantom for evaluating gated blood pool data acquisition/processing systems at heart rates above 400/min

机译:一种廉价的幻象,用于以高于400 / min的心率评估门控血池数据采集/处理系统

获取原文

摘要

Equilibrium gated blood pool imaging of the heart is a common diagnostic procedure for visualizing cardiac function in human subjects. Recently, this procedure has been modified to evaluate cardiac function in mice. However, the high heart rates encountered in these animals (often greater than 400 beats/min) can confound R-wave trigger devices, acquisition systems and image processing software containing default conditions tailored specifically to the lower heart rates of human subjects. In order to determine whether data acquisition and processing components of a commercial or self-generated gated blood pool imaging procedure are performing properly, input of known timing and imaging signals that mimic those generated during high heart rate gated blood pool imaging is required. Here, the authors describe an inexpensive phantom that is suitable for initial evaluation of an unknown system or for ongoing QC of a previously verified system.
机译:心脏的平衡门控血池成像是可视化人类受试者心脏功能的常见诊断程序。最近,已经对该程序进行了修改,以评估小鼠的心脏功能。但是,在这些动物中遇到的高心率(通常大于400次/分钟)会混淆R波触发设备,采集系统和图像处理软件,其中包含专门针对人类受试者较低心率量身定制的默认条件。为了确定商业或自行生成的门控血池成像程序的数据采集和处理组件是否正常执行,需要输入模仿高心率门控血池成像期间生成的信号的已知时序和成像信号。在这里,作者描述了一种廉价的幻影,该幻影适用于未知系统的初始评估或先前验证的系统的持续质量控制。