Correlation of nuclear spectrometer performance with uniformity and resistivity in cadmium zinc telluride

机译:碲化镉锌中核能谱仪性能与均匀度和电阻率的关系

获取原文

摘要

We have used low-temperature photoluminescence (PL) spectroscopy and room-temperature photoluminescence mapping as a measure of composition variation in cadmium zinc telluride grown by high-pressure Bridgman. We have correlated the uniformity, as measured by the line width of the bound exciton peak in the low-temperature spectrum and the degree of variation in the peak position in the room-temperature map, with the peak-to-valley ratio of the 59.5 keV photopeak in the pulse-height spectrum of /sup 241/Am. For detectors having active areas of 10-30 mm/sup 2/, the PL measurements can be combined with resistivity measurements to give a strong predictor of detector performance. For arrays of 1 mm/sup 2/ detectors, the correlation between these material parameters and detector performance is much weaker. Our measurements show that material parameters and detector performance can vary substantially between points on a 1 cm sample.
机译:我们已经使用低温光致发光(PL)光谱和室温光致发光作图来测量由高压Bridgman生长的碲化镉锌中的成分变化。我们已将均匀性与峰谷比59.5进行了关联,该一致性是通过低温光谱中结合的激子峰的线宽和室温图中峰位置的变化程度来衡量的/ sup 241 / Am的脉冲高度谱中的keV光电峰。对于具有10-30 mm / sup 2 /的有效面积的检测器,可以将PL测量值与电阻率测量值结合使用,以提供对检测器性能的有力预测指标。对于1 mm / sup 2 /探测器的阵列,这些材料参数和探测器性能之间的相关性要弱得多。我们的测量结果表明,材料参数和检测器性能可能会在1 cm样品上的各个点之间发生很大变化。