Left ventricular (LV) pressure can be derived noninvasively from the continuous wave Doppler spectra of a mitral regurgitation using the simplified Bernoulli equation. The peak of its first derivative (dP/dt) reflects LV systolic function. Presently, it can be computed on profiles digitised tediously by hand tracings. This paper describes an evaluation of peak dP/dt based on the automatic determination of the contour of regurgitant velocity profiles after digitisation of raw Doppler signals with an A/D board and a PC. This method has been validated in vitro on a pulse duplicator in which a leaking mitral prosthesis was studied under several hemodynamic conditions. The Doppler derived peak dP/dt was well correlated (r/sup 2/=0.9, SEE=251 mmHg) with the measured one. Also recordings on patients showed a good correlation with the mean (1-3 m/s) dP/dt (r/sup 2/=0.7. SEE=365 mmHg), showing a systematic underestimation (199/spl plusmn/176 mmHg). In conclusion, our method is an easy and reliable tool to measure peak LV dP/dt.
机译:可以使用简化的伯努利方程从二尖瓣关闭不全的连续波多普勒频谱中无创地得出左心室(LV)压力。其一阶导数的峰值(dP / dt)反映了LV的收缩功能。目前,它可以在通过手工跟踪将数字化的乏味的轮廓上进行计算。本文描述了在使用A / D板和PC对原始多普勒信号进行数字化后,基于对反流速度分布轮廓的自动确定,对峰值dP / dt的评估。该方法已经在脉冲复制器上进行了体外验证,该脉冲复制器在几种血液动力学条件下研究了泄漏的二尖瓣假体。多普勒推导的峰值dP / dt与所测得的峰高度相关(r / sup 2 / = 0.9,SEE = 251 mmHg)。患者的记录也显示与平均(1-3 m / s)dP / dt(r / sup 2 / = 0.7。SEE = 365 mmHg)良好相关,显示出系统的低估(199 / spl plusmn / 176 mmHg) 。总之,我们的方法是测量峰值LV dP / dt的简便可靠的工具。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号