Angiographic and Doppler ultrasound comparative measurements of mean coronary blood flow

机译:血管造影和多普勒超声对比测量平均冠状动脉血流量

获取原文

摘要

Using results of a computer simulation of the propagation of contrast medium in coronary arteries the authors could improve their angiographic method for measurement of mean coronary blood flow in ml/min. This method is based on 3D-reconstruction of the coronary artery tree from biplane coronary angiograms and computation of the contrast medium distribution along the main branches one or two cardiac cycles after onset of a constant rate injection. Flow values obtained in patients with injection rates above and below peak coronary flow were compared. As expected from the simulation, flow values obtained with the low injection rates were systematically lower than the corresponding ones obtained with the high injection rates. Comparative flow measurements performed with a Doppler "FloWire" system were in good agreement with those obtained angiographically with strong injections.
机译:利用造影剂在冠状动脉中传播的计算机模拟结果,作者可以改进他们的血管造影方法,以ml / min为单位测量平均冠状动脉血流量。此方法基于从双平面冠状动脉血管造影照片对冠状动脉树进行3D重建,并计算出在恒定速率注射后一或两个心动周期沿主要分支的造影剂分布。比较注射速率高于和低于峰值冠脉流量的患者获得的流量值。从仿真中可以预期,低喷射速率获得的流量值系统地低于高喷射速率获得的相应值。用多普勒“ FloWire”系统进行的比较流量测量与通过强力注射通过血管造影获得的流量非常吻合。