首页> 外文会议> >Voltage-current characteristics of an AC arc for different electrode separations with application to electrosurgery II
【24h】

Voltage-current characteristics of an AC arc for different electrode separations with application to electrosurgery II

机译:不同电极间距的交流电弧的电压-电流特性在电外科中的应用II

获取原文

摘要

For pt. I, see ibid., p.35-6 (1993). A study of the effect of linear electrode displacement in various irrigant solutions commonly used during transurethral electrosurgery is discussed. Frequency and electrode type were held constant. A bipolar configuration was used. Results indicated that solutions such as glycine tap water, distilled water and sterile water displayed a capacitive response to the AC applied signal. Electrodes immersed in glycine, distilled water, and sterile water showed no arcing. A complete understanding of the effects of electrode separation and other parameters such as irrigant solutions may help engineers to design hazard free electrosurgical apparatus.
机译:对于pt。我见同上,第35-6页(1993)。讨论了在经尿道电外科手术中常用的各种冲洗液中线性电极位移的影响的研究。频率和电极类型保持恒定。使用了双极配置。结果表明,诸如甘氨酸自来水,蒸馏水和无菌水之类的溶液对交流施加的信号表现出电容性响应。浸入甘氨酸,蒸馏水和无菌水中的电极无电弧。对电极分离和其他参数(例如冲洗液)的影响的全面了解可以帮助工程师设计无害的电外科设备。