The effect of varying filter techniques on late potential parameters of signal averaged electrocardiographic records

机译:各种滤波技术对平均心电图记录后期电位参数的影响

获取原文

摘要

Signal average electrocardiographic parameters were obtained in 54 patients and 128 normal subjects using two filtering techniques and lead systems. Filtered QRS duration (TQRS), root mean voltage in the terminal filtered QRS (RMS), and terminal QRS duration under 40 microvolts (LAD) were compared. TQRS was significantly different for all subjects. RMS values were different for normal subjects and EPS positive subjects. It is concluded that variation in filtering can modify results and that standards appropriate for the filter must be employed clinically.
机译:使用两种滤波技术和导联系统,在54位患者和128位正常受试者中获得信号平均心电图参数。比较了滤波后的QRS持续时间(TQRS),终端滤波后的QRS的均方根电压(RMS)和40微伏以下的终端QRS持续时间(LAD)。对于所有受试者,TQRS均显着不同。正常受试者和EPS阳性受试者的RMS值不同。结论是,过滤器的变化会改变结果,并且必须在临床上采用适合于过滤器的标准液。