Penetration and mass loss rates with different energy levels in the human meniscus for the holmium:YAG laser

机译:men:YAG激光在人类半月板中不同能级的穿透率和质量损失率

获取原文
获取原文并翻译 | 示例

摘要

Abstract: Using a pulsed Holmium:YAG laser, wavelength 2.1 $mu@m, the ablation rate and thermal effect were measured on human meniscal cartilage. Holding the pulsewidth and repetition rate constant, fluences per pulse were varied between 167 - 927 J/cm$+2$//pulse for the penetration rate and between 1.7 - 100 J/cm$+2$//pulse for the mass loss rate. The fastest ablation rate was found at 927 J/cm$+2$//pulse for the penetration rate and at 100 J/cm$+2$//pulse for the mass loss rate. The thresholds were 53 J/cm$+2$//pulse and 1.36 J/cm$+2$//pulse, respectively. The increases in ablation rate were directly proportional to increases in fluences for both methods. Histological examination showed the average lateral thermal change to be 400 - 500 $mu@m and 400 - 460 $mu@m, respectively, with no relation to level of fluence. !0
机译:摘要:使用脉冲2.1:YAG激光,波长为2.1μm@m,测量了对人半月板软骨的消融率和热效应。在保持脉冲宽度和重复率不变的情况下,对于穿透率,每个脉冲的通量在167-927 J / cm $ + 2 $ //脉冲之间变化,对于质量损失,在1.7-100 J / cm $ + 2 $ //脉冲之间变化率。最快的烧蚀速率为927 J / cm $ + 2 $ // pulse的穿透速率和100 J / cm $ + 2 $ // pulse的质量损失速率。阈值分别为53 J / cm $ + 2 $ // pulse和1.36 J / cm $ + 2 $ // pulse。两种方法的消融率增加均与注量的增加成正比。组织学检查显示平均横向热变化分别为400-500 $ mu @ m和400-460 $ mu @ m,与通量水平无关。 !0