308-nm excimer laser ablation of human cartilage

机译:308 nm准分子激光消融软骨

获取原文
获取原文并翻译 | 示例

摘要

Abstract: The XeCl excimer laser was investigated as an ablating tool for human fibrocartilage and hyaline cartilage. Quantitative measurements were made of tissue ablation rates as a function of fluence in meniscal fibrocartilage and articular hyaline cartilage. A force of 1.47 Newtons was applied to an 800 $mu@m fiber with the laser delivering a range of fluences (40 to 190 mj/mm$+2$/) firing at a frequency of 5 Hz. To assess the effect of repetition rate on ablation rate, a set of measurements was made at a constant fluence of 60 mj/mm$+2$/, with the repetition rate varying from 10 to 40 Hz. Histologic and morphometric analysis was performed using light microscopy. The results of these studies revealed that the ablation rate was directly proportional to fluence over the range tested. Fibrocartilage was ablated at a rate 2.56 times faster than hyaline cartilage at the maximum fluence tested. Repetition rate had no effect on the penetration per pulse. Adjacent tissue damage was noted to be minimal (10 - 70 $mu@m). !6
机译:摘要:研究了XeCl准分子激光作为人类纤维软骨和透明软骨的消融工具。定量测量组织消融率与半月板纤维软骨和关节透明软骨的通量的关系。在800μm的光纤上施加了1.47牛顿的力,该激光器以5 Hz的频率发射了一系列的注量(40到190 mj / mm $ + 2 $ /)。为了评估重复率对消融率的影响,在恒定通量为60 mj / mm $ + 2 $ /的情况下进行了一组测量,重复率从10到40 Hz不等。使用光学显微镜进行组织学和形态计量学分析。这些研究的结果表明,在测试范围内,消融率与注量成正比。在测试的最大通量下,纤维软骨的切除速度比透明软骨快2.56倍。重复率对每个脉冲的穿透率没有影响。相邻的组织损伤被认为是最小的(10-70μm@ m)。 !6