Clinical Optimization of Filters in Direct a-Se FFDM (Full Field Digital Mammography) System

机译:直接a-Se FFDM(全场数字乳腺摄影)系统中滤镜的临床优化

获取原文
获取原文并翻译 | 示例

摘要

We evaluated three combinations of filters (Mo/Mo, Mo/ Rh, and W /Rh) in direct a-Se FFDM system to optimize radiation dose clinically. We measured CNR (Contrast to Noise Ratio) as physical characteristics changing radiation dose and phantom thickness in clinical range. In 20, 30, 40, and 50mm PMMA phantoms, Mo / Mo showed the best performances. On the other hand, in 60 and 70 mm, W/Rh 30kV showed best performance. In addition, in 40 and 50mm PMMA phantoms, W/Rh 30kV showed the second best performance. In direct a-Se FFDM system, W/Rh was valuable in minimizing radiation dose.
机译:我们在直接a-Se FFDM系统中评估了过滤器的三种组合(Mo / Mo,Mo / Rh和W / Rh)以优化临床放射剂量。我们测量了CNR(对比度与噪声比),作为在临床范围内改变辐射剂量和幻影厚度的物理特性。在20、30、40和50mm PMMA幻像中,Mo / Mo表现出最佳性能。另一方面,在60和70 mm时,W / Rh 30kV表现出最佳性能。此外,在40mm和50mm PMMA幻像中,W / Rh 30kV表现出第二好的性能。在直接a-Se FFDM系统中,W / Rh在最小化辐射剂量方面很有价值。