首页> 外文会议>International cryogenic engineering conference 1998(ICEC17) >Cryogenic Thermometer Calibration System using a Helium Cooling Loop and a Temperature Controller
【24h】

Cryogenic Thermometer Calibration System using a Helium Cooling Loop and a Temperature Controller

机译:使用氦气冷却回路和温度控制器的低温温度计校准系统

获取原文
获取原文并翻译 | 示例

摘要

The IPN-Orsay and CERN are designing in close collaboration a fully automated cryogenic thermometer calibration facility which will calibrate in 3 years 10,000 cryogenic thermometers required for the Large Hadron Collider (LHC) operation. A reduced-scale model of the calibration facility has been developed, which enables the calibration of ten thermometers by comparison with two Rhodium-Iron standard thermometers in the 1.8 K to 300 K temperature range under vacuum conditions. The particular design, based on a helium cooling loop and an electrical temperature controller, gives good dynamic performances. This paper describes the experimental set-up and the data acquisition system. Results of experimental runs are also presented along with the estimated global accuracy for the calibration.
机译:IPN-Orsay和CERN正在紧密合作设计一个全自动的低温温度计校准设备,该设备将在3年内校准大型强子对撞机(LHC)操作所需的10,000个低温温度计。已经开发出一种缩小规模的校准设备模型,通过与两个在真空条件下在1.8 K至300 K温度范围内的Rhodium-Iron标准温度计进行比较,可以校准十个温度计。基于氦冷却回路和电气温度控制器的特殊设计具有良好的动态性能。本文介绍了实验设置和数据采集系统。还提供了实验运行的结果以及估计的整体校准精度。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号