摘要

The purpose of the study was to apply European Commission (EC) Reference Dose Levels (RDL) in ComputedrnTomography (CT) examinations at Onassis Cardiac Surgery Center (OCSC). These are weighted CT Dose Indexrn(CTDIw) for a single slice and Dose-Length Product (DLP) for a complete examination. During the period 1998-rn1999, the total number of CT examinations, every type of CT examination, patient related data and technicalrnparameters of the examinations were recorded. The most frequent examinations were chosen for investigationrnwhich were the head, chest, abdomen and pelvis. CTDI measurements were performed and CTDIw and DLP wererncalculated. Third Quartile values of CTDIw were chosen to be 43mGy for head, 8mGy for chest, and 22mGy forrnabdomen and pelvis examinations. Third Quartile values of DLP were chosen to be 740mGycm for head,rn370mGycm for chest, 490mGycm for abdomen and 420mGycm for pelvis examination. Results confirm thatrnOCSC follows successfully the proposed RDL for the head, chest, abdomen and pelvis examinations in terms ofrnradiation dose.
机译:该研究的目的是在欧纳西斯心脏外科中心(OCSC)的计算机断层扫描(CT)检查中应用欧洲委员会(EC)参考剂量水平(RDL)。这些是单个切片的加权CT剂量指数(CTDIw)和完整检查的剂量长度乘积(DLP)。在1998年至1999年期间,记录了CT检查的总数,每种类型的CT检查,患者相关数据和检查的技术参数。选择用于检查的最频繁的检查是头部,胸部,腹部和骨盆。进行CTDI测量并计算CTDIw和DLP。 CTDIw的第三四分位数选择为:头部43mGy,胸部8mGy,腹部和骨盆检查22mGy。 DLP的第三四分位数选择为:头部740mGycm,胸部rn370mGycm,腹部490mGycm和骨盆检查420mGycm。结果证实,rnOCSC已成功遵循建议的RDL进行了头部,胸部,腹部和骨盆的放射剂量检查。