首页> 外文会议>IEEE Nuclear Science Symposium;Medical Imaging Conference >New Gd Scintillation Glass Fiber Arrays for High Spatial Resolution Cold Neutron Imaging
【24h】

New Gd Scintillation Glass Fiber Arrays for High Spatial Resolution Cold Neutron Imaging

机译:用于高空间分辨率冷中子成像的新型Gd闪烁玻璃纤维阵列

获取原文

摘要

The new Gd Scintillation glass fiber arrays as a key component of cold neutron detector composed of 10-μm diameter Gd glass fibers with a center-to-center distance of 12 μm have been invented and fabricated. A cooled Andor iKon-LDZ936 charge-coupled device(CCD) coupled with a Nikon Nikkor 135-mm lens was used to read out the scintillation light from the array. It was found that the detection system has a spatial resolution of about 50 μm. The achieved spatial resolution is mainly limited by the low resolution of the optical readout system. To further characterize the spatial resolution of these arrays, a higher resolution optical readout system is desired.
机译:新型的Gd闪烁玻璃纤维阵列作为冷中子探测器的关键部件,由直径10μm的Gd玻璃纤维组成,中心距为12μm。使用冷却的Andor iKon-LDZ936电荷耦合器件(CCD)和Nikon Nikkor 135毫米透镜,从阵列中读出闪烁光。发现该检测系统具有约50μm的空间分辨率。所获得的空间分辨率主要受光学读出系统的低分辨率限制。为了进一步表征这些阵列的空间分辨率,需要更高分辨率的光学读出系统。