The Detector Characterization Laboratory at NASA/GSFC has investigated the reciprocity failure characteristics of 1.7μm cut-off HgCdTe devices provided by Teledyne Imaging Sensors to the Hubble Space Telescope (HST) Wide Field Camera 3 (WFC3) project. The reciprocity failure follows a power law behavior over the range of fluxes tested (0.1-10~4 photons/second). The slope of the power law varies among detectors, ranging from ~0.3-1%/dex at 1.0μm, which is much smaller than the ~6%/dex effect observed with the HST NICMOS 2.5μm cut-off detectors. In addition, the reciprocity failure exhibits no wavelength dependence, although only a restricted range of wavelengths (0.85-1.0μm) has been explored to date. Despite its relatively small magnitude, reciprocity failure is nevertheless an important effect in the calibration of WFC3 data, as well as in other applications in which there is a large difference in flux between the photometric standards and the scientific sources of interest.
机译:NASA / GSFC的检测器表征实验室研究了由Teledyne成像传感器提供给哈勃太空望远镜(HST)广角相机3(WFC3)项目的1.7μm截止HgCdTe装置的互易失效特性。互易失效遵循在测试通量范围内(0.1-10〜4光子/秒)的幂律行为。功率定律的斜率在各检测器之间变化,在1.0μm时约为0.3-1%/ dex,远小于使用HST NICMOS2.5μm截止检测器观察到的约6%/ dex效应。此外,尽管迄今为止仅探索了有限的波长范围(0.85-1.0μm),但互易失效不表现出波长依赖性。尽管互易失效幅度较小,但它在WFC3数据的校准以及光度标准和感兴趣的科学来源之间通量差异很大的其他应用中仍然是重要的影响。