首页> 外文会议>Hard x-ray, gamma-ray, and neutron detector physics XII >X-ray detection method for the direct-conversion plasma detector
【24h】

X-ray detection method for the direct-conversion plasma detector

机译:直接转换等离子体检测器的x射线检测方法

获取原文
获取原文并翻译 | 示例

摘要

In this study 2-dimensional plasma simulation code was used to find a new approach of x-ray detection method with PDP-like geometry and condition. A conventional PDP geometry consisted of three electrodes was selected and Ne-Xe composition gas was filled the cell-gap. Depending on incident X-ray energy, the number of charges generated within the cell-gap could be different. For the charge amplification and collection of charges two consecutive ac pulses were applied to scan and address electrodes. The increased charges were collected on the positive-bias address electrode at the rear panel. Two parameters, such as amplitude of collected current and formation delay of collected current, were calculated and compared. The formation delay showed more accurate relationship than the collected current amplitude.
机译:在这项研究中,二维等离子体模拟代码被用于寻找一种具有类似PDP几何形状和条件的X射线检测方法的新方法。选择由三个电极组成的常规PDP几何形状,并在单元间隙中填充Ne-Xe组成气体。根据入射的X射线能量,在细胞间隙内产生的电荷数量可能不同。为了电荷的放大和电荷的收集,将两个连续的交流脉冲施加到扫描和寻址电极。增加的电荷收集在后面板上的正偏置地址电极上。计算并比较了两个参数,例如收集电流的幅度和收集电流的形成延迟。形成延迟显示出比收集的电流幅度更精确的关系。