REMIR is the NIR camera of the automatic REM (Rapid Eye Mount) Telescope located at ESO-La Silla Observatory (Chile) and dedicated to monitor the afterglow of Gamma Ray Burst events. During the last two years, the REMIR camera went through a series of cryogenics problems, due to the bad functioning of the Leybold cryocooler Polar SC7. Since we were unable to reach with Leybold for a diagnosis and a solution for such failures, we were forced to change drastically the cryogenics of REMIR, going from cryocooler to LN2: we adopted an ad-hoc modified Continuous Flow Cryostat, a cryogenics system developed by ESO and extensively used in ESO instrumentation, which main characteristic is that the LN2 vessel is separated from the cryostat, allowing a greater LN2 tank, then really improving the hold time. In this paper we report the details and results of this operation.
机译:REMIR是位于智利ESO-La Silla天文台的自动REM(快速眼挂)望远镜的近红外摄像机,专用于监视伽玛射线暴事件的余辉。在过去的两年中,由于Leybold低温制冷器Polar SC7的不良功能,REMIR相机遇到了一系列低温问题。由于我们无法与Leybold进行联系以提供诊断和解决此类故障的方法,因此我们被迫彻底改变了REMIR的低温技术,从低温冷却器到LN2:我们采用了临时改进的连续流低温恒温器,这是一种低温系统由ESO制造,并广泛用于ESO仪器中,其主要特点是将LN2容器与低温恒温器分开,允许使用更大的LN2储罐,从而真正提高了保持时间。在本文中,我们报告了此操作的详细信息和结果。