【24h】

Results of the International Comparison of Absolute Gravimeters in Walferdange (Luxembourg) of November 2003

机译:2003年11月在Walferdange(卢森堡)的国际重力仪国际比较的结果

获取原文
获取原文并翻译 | 示例

摘要

The results of an international comparison of absolute gravimeters held in Walferdange Luxembourg in November 2003 are presented. The absolute meters agreed with one another with a standard deviation of less than 2 μGal (1 Gal = 1 cm/s~2) (if we exclude one prototype instrument from the analysis). For the first time, the ability of the operators was put to the test. The comparison indicates that the errors due to the operator are less than 1 μGal, i.e. within the observational errors.
机译:介绍了2003年11月在卢森堡Walferdange举行的国际重力仪国际比较的结果。绝对仪表相互之间的标准偏差小于2μGal(1 Gal = 1 cm / s〜2)(如果我们从分析中排除一种原型仪器)。首次对操作员的能力进行了测试。比较表明,由于操作者引起的误差小于1μGal,即在观察误差之内。