Laser-induced fluorescence spectrometer based on solid state tunable color-center laser for heavy metal analysis

机译:基于固态可调色心激光器的激光诱导荧光光谱仪用于重金属分析

获取原文
获取原文并翻译 | 示例

摘要

Abstract: All solid state (SS) laser induced fluorescence (LIF) spectrometer based on tunable LiF:F$-2$/$+$MIN$/ color center (C.C.) MALSAN laser with second and fourth harmonic generation was optimized for trace metallic concentration measurements in water solutions. Two types of heavy metals atoms Sn and Pb were analyzed using inexpensive low temperature types of the propane/butane-air flames. The maximum sensitivity of 10 and 50 ppb with good linearity of fluorescence intensity concentration dependence was obtained for Sn and Pb atoms, respectively, using a monochromator for the fluorescence wavelength selection. When the monochromator was replaced by a set of color optical filters the sensitivity of 0.3 ppb for Pb atoms was obtained. Hence, the sensitivity was increased more than 150 times for more cheap and simple registration set-up. An attempt was made to use more compact super- broadband LiF:F$-2$/$+$MIN$/ C.C. laser to replace the MALSAN laser in LIF spectrometer. Up to now the sensitivity of 1 ppm for Pb with the monochromator based fluorescence registration system was obtained in the latter case. !11
机译:摘要:基于可调谐LiF:F $ -2 $ / $ + $ MIN $ /色心(CC)MALSAN激光的全固态(SS)激光诱导荧光(LIF)光谱仪具有二次和四次谐波生成,适用于痕量金属水溶液中的浓度测量。使用便宜的低温类型的丙烷/丁烷空气火焰分析了两种类型的重金属原子Sn和Pb。使用单色仪进行荧光波长选择,分别对Sn和Pb原子获得了10和50 ppb的最大灵敏度,并具有良好的线性的荧光强度浓度依赖性。用一组彩色滤光片代替单色仪时,对Pb原子的灵敏度为0.3 ppb。因此,为了更便宜和更简单的注册设置,灵敏度提高了150倍以上。试图使用更紧凑的超宽带LiF:F $ -2 $ / $ + $ MIN $ /C.C。激光取代了LIF光谱仪中的MALSAN激光。到目前为止,在后一种情况下,使用基于单色仪的荧光配准系统对Pb的灵敏度为1 ppm。 !11

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号