A digital light projector is implemented as an integrated illumination source and scanning element in a confocal non-mydriatic retinal camera, the Digital Light Ophthalmoscope (DLO). To simulate scanning, a series of illumination lines are rapidly projected on the retina. The backscattered light is imaged onto a 2-dimensional rolling shutter CMOS sensor. By temporally and spatially overlapping the illumination lines with the rolling shutter, confocal imaging is achieved. This approach enables a low cost, flexible, and robust design with a small footprint. The 3rd generation DLO technical design is presented, using a DLP LightCrafter 4500 and USB3.0 CMOS sensor. Specific improvements over previous work include the use of yellow illumination, filtered from the broad green LED spectrum, to obtain strong blood absorption and high contrast images while reducing pupil constriction and patient discomfort.
机译:数字光投影仪被实现为共焦非散瞳视网膜相机数字光检眼镜(DLO)中的集成照明源和扫描元件。为了模拟扫描,将一系列照明线快速投射到视网膜上。反向散射的光被成像到二维滚动快门CMOS传感器上。通过使照明线在时间和空间上与卷帘快门重叠,可以实现共聚焦成像。这种方法可实现低成本,灵活且坚固的设计,并且占地面积小。介绍了使用DLP LightCrafter 4500和USB3.0 CMOS传感器的第三代DLO技术设计。与以前的工作相比,具体的改进包括使用从宽泛的绿色LED光谱中滤除的黄色照明,以获得强大的血液吸收和高对比度图像,同时减少瞳孔缩小和患者不适。