Degradation of Type PR and R Thermocouples during Long-term Creep Test

机译:长期蠕变测试期间PR和R型热电偶的性能下降

获取原文
获取原文并翻译 | 示例

摘要

Controlling of accurate testing temperature is most important for reliable creep test. However, degradation of thermocouples occurs during long-term service for creep test at the elevated temperatures. Thermocouples have been calibrated prior to creep test and thermal electromotive force of those has been measured after service. Magnitude of drift was very small for the thermocouples used at 550℃ and below of that, however, it increased with increasing temperature. Degradation of thermocouples has been sufficiently recovered by repair treatment that is annealing and re-welding. Thermocouples used for the creep test of NRIM/NIMS Creep Data Sheets have been conducted to repair treatment after every creep test regardless of testing time, and a drift of Type R thermocouples has been maintained within ± 4℃ for 100,000h at 650℃.
机译:对于可靠的蠕变测试,控制精确的测试温度至关重要。但是,在高温下进行长期蠕变测试时,热电偶会发生降解。热电偶已经在蠕变测试之前进行了校准,并且在使用后已经测量了它们的热电动势。 550℃及以下温度下使用的热电偶的漂移幅度很小,但随温度升高而增加。通过进行退火和重新焊接的修复处理,热电偶的降解已得到充分恢复。在每次蠕变测试之后,无论测试时间如何,都已进行了NRIM / NIMS蠕变数据表蠕变测试所用的热电偶进行修复处理,并且在650℃下,R型热电偶的漂移保持在±4℃100,000h之内。