首页> 外文会议>Conference on Three-Dimensional and Multidimensional Microscopy: Image Acquisition and Processing XI; 20040127-20040129; San Jose,CA; US >Optimization of Spinning Disk Confocal Microscopy: Synchronization with the Ultra-Sensitive EMCCD
【24h】

Optimization of Spinning Disk Confocal Microscopy: Synchronization with the Ultra-Sensitive EMCCD

机译:转盘共聚焦显微镜的优化:与超灵敏的EMCD同步

获取原文
获取原文并翻译 | 示例

摘要

The advent of Electron Multiplying Charge Coupled Device (EMCCD) technology and it's ability to overcome previous hurdles in low-light fluorescence microscopy, such as phototoxicity to live cells, photobleaching of fluorophores and exposure time restrictions, has resulted in a significant resurgence of interest in use of confocal spinning disk techniques for live cell microscopy. Here provide an understanding of, and technical solutions to, the issues of synchronization that have previously marred the coupling of fast CCD camera technology to confocal spinning disk arrangements. We examine the challenges arising from both old and new models of the Nipkow spinning disk confocal unit and suggest solutions throughout based on a sound comprehension of both (a) relative scan/exposure times; (b) relative orientation of the coupled devices; (c) optimisation of EMCCD clocking parameters.
机译:电子倍增电荷耦合器件(EMCCD)技术的出现以及克服弱光荧光显微镜以前的障碍的能力,例如对活细胞的光毒性,荧光团的光漂白和暴露时间的限制,引起了人们的极大兴趣使用共聚焦旋转盘技术进行活细胞显微镜检查。这里提供对同步问题的理解和技术解决方案,这些问题以前使快速CCD摄像头技术与共聚焦旋转磁盘装置的耦合受到损害。我们研究了Nipkow旋转圆盘共聚焦装置的新旧模型带来的挑战,并基于对以下两个方面的合理理解,提出了解决方案:(a)相对扫描/曝光时间; (b)耦合设备的相对方向; (c)优化EMCCD时钟参数。