Measurement Of Mucosal Capillary Hemoglobin Oxygen Saturation in the Colon By Reflectance Spectrophotometry

机译:反射分光光度法测定结肠黏膜毛细血管血红蛋白氧饱和度

获取原文
获取原文并翻译 | 示例

摘要

Advances in optical and computer technology have enabled the development of a device that utilizes white-light reflectance spectrophotometry to measure capillary hemoglobin saturation in intestinal mucosa during colonoscopy. Studies were performed using the colon oximeter in anesthetized animals and patients undergoing colonoscopy. Mucosal hemoglobin saturation in the normal colon (mean +/- S.D.) is 72% +/- 3.5%. In an animal model, ischemia via arterial ligation and hypoxemia via hypoxic ventilation each result in a decrease of over 40% in the mucosal saturation. In human patients with colon polyps, ischemia induced by epinephrine injection, stalk ligation using a loop, or clipping of the polyp stalk each result in a decrease of over 40% in the mucosal saturation (p<0.02). In contrast, saline injection does not decrease the mucosal saturation (p=N.S.). A patient who previously underwent partial colectomy with sacrifice of the inferior mesenteric artery had a saturation of 55% in the remaining sigmoid colon, with normal values in the superior mesenteric artery territory (p<0.05). A novel device for measuring capillary hemoglobin saturation in intestinal mucosa during colonoscopy is capable of providing reproducible measurements in normal patients and clearly detects dramatic decreases in saturation with ischemic and hypoxic insults.
机译:光学和计算机技术的进步使得能够开发一种利用白光反射分光光度法在结肠镜检查过程中测量肠粘膜中毛细血管血红蛋白饱和度的设备。使用结肠血氧仪对麻醉的动物和接受结肠镜检查的患者进行了研究。正常结肠的粘膜血红蛋白饱和度(平均+/- S.D.)为72%+/- 3.5%。在动物模型中,通过动脉结扎引起的局部缺血和通过低氧通气引起的低氧血症均导致粘膜饱和度降低40%以上。在患有结肠息肉的人类患者中,肾上腺素注射,使用环套扎茎杆或切断息肉茎杆引起的局部缺血均导致粘膜饱和度降低40%以上(p <0.02)。相反,盐水注射不会降低粘膜饱和度(p = N.S。)。先前因牺牲肠系膜下动脉而接受部分结肠切除术的患者,其余乙状结肠的饱和度为55%,而在肠系膜上动脉区的正常值(p <0.05)。一种用于在结肠镜检查期间测量肠粘膜中毛细血管血红蛋白饱和度的新颖设备,能够为正常患者提供可重复的测量结果,并清楚地检测出缺血和缺氧性损伤导致的饱和度急剧下降。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号