摘要

Miniball is a dedicated Ge detector array which has been developed for the investigation of rare γ decays at the new radioactive beam facility REX-ISOLDE at CERN. The array is optimised for high full-energy peak efficiency and for high granularity needed to perform Doppler corrections of γ-rays emitted by fast moving nuclei. Miniball will finally consist of 40 six-fold segmented, encapsulated detectors which are clustered in eight cryostats with three detectors each and four cryostats with four detectors, respectively. It is shown that from an analysis of the pulse shapes and of the amplitudes of the mirror charges in the adjacent segments the effective granularity of Miniball can be enhanced from 240 to ≈ 4000. The properties of Miniball are compiled on the basis of experimental data. Examples of the first data measured with Miniball are presented.
机译:Miniball是专用的Ge探测器阵列,已开发用于研究CERN新的放射性束设施REX-ISOLDE中罕见的γ衰变。该阵列经过优化,可实现高全能峰值效率和执行快速移动核发射的γ射线多普勒校正所需的高粒度。微型球最终将由40个六段分段封装的探测器组成,这些探测器聚集在八个具有三个探测器的低温恒温器和四个分别具有四个探测器的低温恒温器中。结果表明,通过对相邻段中的脉冲形状和镜面电荷的幅度进行分析,可以将Miniball的有效粒度从240提高到≈4000。Miniball的性能是根据实验数据进行编译的。给出了用Miniball测量的第一批数据的示例。