Field screening of petroleum hydrocarbons using synchronous scanning fluorescence spectrometry

机译:同步扫描荧光光谱法现场筛选石油烃

获取原文
获取原文并翻译 | 示例

摘要

The U.S. Environmental Protection Agency (EPA), National Exposure Research Laboratory (NERL), Characterization Research Division in Las Vegas, Nevada (CRD-LV) is continuing to evaluate the perofrmance of a prototype portable scanning ultravioelt-visible fluorescence instrument as a field scrrening toll. This field ortable scanning spectrofluorometer (FPSS) can rapidly detect the type and relative amount of hydrocarbon contamination in soil and water samples. A recent on site spectrofluorometric analysis of soil and water samples was conducted at an oil refinery in Billings, Montana. Soil and water samples (100 g for soil; 40 mL VOA vial for water) were collected and extracted with iso-ocatane. The extractants were rapidly scanned from 210 nm to 480 nm in the synchronous fluorescence mode. A wavelength offset between the excitation and emission monochromators of 10 nanometers (nm) provided simplified unique spectral fingerprint scans with fluorescence intensity proportional to contaminant concentration. Spectra from environmental samples were compared with site-specific reference standards, including crude oil, fuel oil No. 2 and fuel oil No. 6. The results of the on-site field analyses gave site investigators immediate data that showed the type of hydrocarbon contamination and the relative amounts. Immediate availability of this type of information allows a field investigator to take prompt corrective action as necessary.
机译:美国环境保护局(EPA),内华达州拉斯维加斯特征研究部国家暴露研究实验室(NERL)(CRD-LV)继续评估便携式便携式扫描紫外可见荧光仪器的性能,以作为现场测试。收费。这种现场稳定的扫描荧光分光光度计(FPSS)可以快速检测土壤和水样中碳氢化合物污染的类型和相对量。最近在蒙大拿州比林斯的一家炼油厂对土壤和水样进行了现场荧光光谱分析。收集土壤和水样本(土壤100克;水40 mL VOA小瓶),并用异卡他汀萃取。在同步荧光模式下,将提取剂从210 nm快速扫描到480 nm。激发和发射单色仪之间的10纳米(nm)波长偏移提供了简化的独特光谱指纹扫描,其荧光强度与污染物浓度成正比。将来自环境样品的光谱与特定地点的参考标准(包括原油,2号燃料油和6号燃料)进行了比较。现场现场分析的结果为现场调查人员提供了即时数据,显示了碳氢化合物污染的类型和相对金额。立即获得此类信息可以使现场调查人员根据需要迅速采取纠正措施。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号