A parallel confocal method including an optical diffractive element, a Dammann grating, is presented. Dammann gratings manufactured with microfabrication technologies have the ability to generate several beams with identical intensity. An optical confocal system combining with a Dammann grating will form multi-channel confocal optical structures, so that it can achieve a planar image instantly without scanning. This parallel confocal fluorescence detection system has the merits of discarding scanning part, compact structure, real time imaging and detection and high S/N ratio, therefore it is suitable for detection of dye-labelled DNA array chips and for other biomedical applications. Structural considerations and design of a parallel confocal fluorescence setup and experimental results of detecting a dye-labeled DNA array chip are presented in this paper.
机译:提出了一种包括光衍射元件,达曼光栅的平行共焦方法。用微细加工技术制造的达曼光栅具有产生几束相同强度的光束的能力。与Dammann光栅结合的光学共焦系统将形成多通道共焦光学结构,因此无需扫描即可立即获得平面图像。该平行共聚焦荧光检测系统具有扫描部分省去,结构紧凑,实时成像和检测,信噪比高等优点,因此适用于染料标记DNA阵列芯片的检测以及其他生物医学应用。本文介绍了平行共焦荧光设置的结构考虑和设计,以及检测染料标记的DNA阵列芯片的实验结果。