Effects of deep breathing on blood pressure measurement in healthy subjects

机译:深呼吸对健康受试者血压测量的影响

获取原文
获取原文并翻译 | 示例

摘要

The aim of this study was to investigate the effects of deep breathing on blood pressure parameters: systolic (SBP), diastolic (DBP) and pulse pressure, mean pulse rate (MPR), and the amplitude of Korotkoff sounds. Manual SBP and DBP were measured from 20 healthy subjects (age: 44±13 years) during normal (NB) and deep (DB) breathing. A chest magnetometer was used to acquire the respiratory depth, and an electronic stethoscope for the Korotkoff sound. All data were sampled at 2 kHz. MPR was derived from the oscillometric pulse waveform. For each oscillometric pulse, time-frequency analysis of the Korotkoff sound located the time of the peak energy (Tp), and Korotkoff sounds amplitude was defined as the maximum peak-to-peak amplitude in the window TP±50ms. The respiratory rate was calculated from both the magnetometer (fR) and Korotkoff amplitude series (fK) for the period between the manually measured SBP and DBP, with a power spectral density resolution of 0.016 Hz in the physiological respiratory frequency range at rest. DB compared to NB increased respiratory regularity (quantified by spectral concentration) by 9±10% (mean±SD, p<0.005), and decreased fR from 0.24±0.06 to 0.20±0.05 Hz (p<0.0005). SBP and pulse pressure also decreased, from 116±12 to 113±11 mmHg (p<0.005) and 41±9 to 37±9 mmHg (p<0.01) respectively, and MPR increased from 64±8 to 68±9 pulse/min (p<0.0005). In 70% of recordings fR differed from fK by 0.05 Hz or less. The results show an effect of DB on blood pressure, and the presence of a modulating effect of respiration on the Korotkoff sound.
机译:这项研究的目的是研究深呼吸对血压参数的影响:收缩压(SBP),舒张压(DBP)和脉压,平均脉搏率(MPR)以及Korotkoff声音的振幅。在正常(NB)和深度(DB)呼吸期间,对20名健康受试者(年龄:44±13岁)进行了手动SBP和DBP测量。使用胸部磁力计获取呼吸深度,并使用电子听诊器获取Korotkoff声音。所有数据均以2 kHz采样。 MPR是从示波脉冲波形得出的。对于每个示波脉冲,对Korotkoff声音的时频分析确定了峰值能量(Tp)的时间,并且Korotkoff声音振幅被定义为窗口T P ±50ms。从磁力计(f R )和Korotkoff振幅序列(f K )两者计算出人工呼吸SBP和DBP之间的呼吸频率,并得出功率谱在静止的生理呼吸频率范围内,密度分辨率为0.016 Hz。与NB相比,DB的呼吸规律性(按频谱浓度量化)增加了9±10%(平均值±SD,p <0.005),并且f R 从0.24±0.06降低至0.20±0.05 Hz(p <0.0005)。 SBP和脉压也分别从116±12到113±11 mmHg(p <0.005)和41±9到37±9 mmHg(p <0.01)降低,MPR从64±8到68±9pulse /最小值(p <0.0005)。在70%的录音中,f R 与f K 相差0.05 Hz或更小。结果显示了DB对血压的影响,以及呼吸对Korotkoff声音的调节作用。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号