STRESS ANALYSIS OF THE POLYETHYLENE COMPONENT IN TOTAL ANKLE ARTHROPLASTY: EFFECT OF THICKNESS

机译:全踝关节置换术中聚乙烯组分的应力分析:厚度的影响

获取原文
获取原文并翻译 | 示例

摘要

A three-dimensional finite element model was constructed of the distal tibia and fibula and a semi-constrained ankle prosthesis (Agility~(TM) system). Contact elements were used at the interface between the talar component and the polyethylene liner and the proximal tibia and fibular were loaded in the in vertical direction. The minimal thickness of the the polyethylene liner was varied from 3 mm to 8 mm in 1 mm increments. The results showed that the liner contact pressure in the sagittal plane mid-line decreased from 20 MPa to 14 MPa with increasing thickness while the medial edge contact pressure increased from 26 MPa to 30 MPa.
机译:建立了胫骨远端和腓骨和半约束踝关节假体(Agility〜(TM)系统)的三维有限元模型。在距骨组件和聚乙烯内衬之间的界面处使用接触元件,并沿垂直方向加载近端胫骨和腓骨。聚乙烯衬里的最小厚度以1毫米为增量从3毫米变化到8毫米。结果表明,随着厚度的增加,矢状面中线的衬里接触压力从20 MPa降低到14 MPa,而内侧边缘接触压力从26 MPa增加到30 MPa。