Automatic distance measurement of abdominal aorta for ultrasonography-based visceral fat estimation

机译:基于超声检查的内脏脂肪自动测量腹主动脉的距离

获取原文

摘要

Ultrasonography-based visceral fat estimation is a promising method to assess central obesity, which is associated with metabolic syndrome. The key to this method is to measure three types of distance in the ultrasound image. The most important one is the distance from the skin surface to the posterior wall of the abdominal aorta. We present a novel automatic measurement method to calculate this distance using 1D ultrasound signal processing. It is different from the conventional 2D image processing based methods which have high failure rate when the target is blurred or partially imaged. The proposed method identifies the waveforms of the aorta along a group of ultrasound scan lines and a rating mechanism is introduced to choose the best waveform for distance calculation. The robustness and accuracy of the method were evaluated by experiments based on clinical data.
机译:基于超声检查的内脏脂肪估计是一种有前途的评估与代谢综合征相关的中心性肥胖的方法。该方法的关键是测量超声图像中的三种距离。最重要的是从皮肤表面到腹主动脉后壁的距离。我们提出一种新颖的自动测量方法,以使用一维超声信号处理来计算该距离。它与基于常规2D图像处理的方法不同,后者在目标模糊或部分成像时具有较高的失败率。所提出的方法沿着一组超声扫描线识别主动脉的波形,并引入一种评级机制来选择最佳波形进行距离计算。通过基于临床数据的实验评估了该方法的鲁棒性和准确性。