首页> 外文会议>Advances in sensors, signals, visualization, imaging and simulation >A Feasibility Study of Heartbeat Detections from Photoplethysmograms With Fingers Compressed
【24h】

A Feasibility Study of Heartbeat Detections from Photoplethysmograms With Fingers Compressed

机译:用手指压迫从光体积描记图检测心跳的可行性研究

获取原文
获取原文并翻译 | 示例

摘要

In this paper, we study the influence of a pressure applied to the fingers when measuring photoplethysmographic (PPG) signals. We built our own PPG device that acquires the applied finger forces and standard electrocardiographic signals in parallel with the PPG measurements. An experiment was carried out with 8 test subjects whose PPG signals were recorded in 30 8-second intervals with finger forces increasing from 0 to 8.83 N in steps of 0.59 N, and then decreasing back to 0. We developed a method to detect PPG-based heartbeats. The obtained results yield sensitivity, specificity, and precision of the detections of 95.6%, 94.2%, and 95.6%, respectively. We also present the quality of detections versus the forces applied with fingers.
机译:在本文中,我们研究了在测量光电容积描记(PPG)信号时施加到手指的压力的影响。我们构建了自己的PPG设备,该设备在与PPG测量同时获取手指施加的力和标准心电图信号。对8位测试对象进行了实验,每30个8秒间隔记录一次PPG信号,手指力从0增大到8.83 N,步长为0.59 N,然后减小到0。我们开发了一种检测PPG-基于心跳。所得结果的灵敏度,特异性和检测精度分别为95.6%,94.2%和95.6%。我们还介绍了检测质量与手指施加的力。