Patients with traumatic brain injury are difficult to be moved and the patients cannot get prompt treatment, but the real-time detection can be carried out for the intracranial disease. Changes of moisture content in brain can represent the severity of cerebral edema directly. This paper presents an open nuclear magnetic resonance (NMR) system for brain edema detection. It can detect the moisture content in different positions of the brain by measuring NMR signals at different depths in target area. After demonstrating the optimal selections of the main magnet and radio frequency (RF) coil, a printed circuit board (PCB) of bilayer splayed coil and its capacitor matching circuit are made, then an open NMR system is established. Using flour and cosmetic cotton in the magnet target area as samples, the NMR system's reliability and accuracy are proved through experiments. Experimental results show that the system designed can detect moisture content in the object at different positions, and consequently for real-time detection of cerebral edema. Due to the system's portability, miniaturization, and low cost, it can provide experiment support for further application of open NMR instrument in clinical.
机译:颅脑外伤患者难以移动,无法及时治疗,但可以实时检测颅内疾病。脑中水分含量的变化可以直接代表脑水肿的严重程度。本文提出了一种用于脑水肿检测的开放式核磁共振(NMR)系统。它可以通过测量目标区域不同深度的NMR信号来检测大脑不同位置的水分含量。在说明了主磁体和射频线圈的最佳选择之后,制作了双层张开线圈的印刷电路板(PCB)及其电容器匹配电路,然后建立了一个开放式NMR系统。以磁铁目标区域的面粉和化妆棉为样本,通过实验证明了核磁共振系统的可靠性和准确性。实验结果表明,所设计的系统可以检测物体不同位置的水分,从而可以实时检测脑水肿。由于该系统的便携性,小型化和低成本,它可以为开放式NMR仪器在临床上的进一步应用提供实验支持。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号