This paper presents an improved indirect feedback instrumentation amplifier for reading ECG signals. The circuit uses a charge pump module and an integral feedback offset suppression module. The input common mode range (ICMR) is extended, and the electrode offset and baseline drift in the ECG signal are suppressed. The circuit is designed by using 0.18μm CMOS process. The simulation shows that the circuit can operate at the input offset voltage 200mV at the 2V supply voltage. The input common mode range is 0~3.4v, the gain is 46dB, the power consumption is 121.6μW, and the common mode rejection ratio is 124dB.
机译:本文提出了一种改进的间接反馈仪表放大器,用于读取ECG信号。该电路使用电荷泵模块和积分反馈失调抑制模块。扩展了输入共模范围(ICMR),并抑制了ECG信号中的电极偏移和基线漂移。该电路采用0.18μmCMOS工艺设计。仿真表明,该电路可以在2V电源电压下的200mV输入失调电压下工作。输入共模范围为0〜3.4v,增益为46dB,功耗为121.6μW,共模抑制比为124dB。