Evaluation of biocompability of supramolecular calixarene on L-929 by real time cell analysis

机译:通过实时细胞分析评估超分子杯芳烃在L-929上的生物相容性

获取原文
获取原文并翻译 | 示例

摘要

The cytotoxicity test is one of the biological evaluation and screening tests that use tissue cells in vitro to observe the cell growth, reproduction and morphological effects by medical devices. Cytotoxicity is preferred as a pilot project test and an important indicator for toxicity evaluation of medical devices as it is simple, fast, has a high sensitivity and can save animals from toxicity. Three types of cytotoxicity test are stated in the International Organization for Standardization 109993-5: Extract, direct contact and indirect contact tests. The xCELLigence real-time cell analysis system shows a significant potential in regards to cytotoxicity in recent years. The proliferation of L-929 were observed in the Real-Time Cell Analysis (RTCA) Systems. Our research directs that calixarene shows no cytotoxicity at a concentration range of 25 to 100 μg/mL on L929 cells. There were no significant differences between the L929 cell indexes of the control and all testing groups at 24, 48 and 72 h.
机译:细胞毒性测试是生物学评估和筛选测试之一,该测试使用体外组织细胞观察医疗器械对细胞生长,繁殖和形态的影响。细胞毒性是简单的,快速的,具有高灵敏度并可以使动物免于毒性的试验,因此优选作为试验项目测试和医疗器械毒性评估的重要指标。国际标准化组织109993-5规定了三种类型的细胞毒性测试:提取,直接接触和间接接触测试。近年来,xCELLigence实时细胞分析系统在细胞毒性方面显示出巨大潜力。在实时细胞分析(RTCA)系统中观察到L-929的增殖。我们的研究表明,杯状芳烃在L929细胞上的浓度范围为25至100μg/ mL时无细胞毒性。在24、48和72小时,对照组和所有测试组的L929细胞指数之间没有显着差异。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号