There is an increasing need to develop low power medical devices for healthcare monitoring applications. Recently advancement in microelectronics, micro electro mechanical systems (MEMS), telecommunications and micro biomedical sensors have enabled both development and design of such devices. Different solutions have been developed such as wireless sensor networks (WSN), intelligent habitat systems and so on. In this paper, we focus on the actimery concept. The aim is the design of RF transmitter architecture to ensure the real time data transmission. The proposed architecture is based on the IEEE 802.15.4a standard and it operates in the mandatory channel 3 of the low band. The transmitter is expected to working at a data rate of 0.850 Kbps and using BPSK modulation. The simulated TX has peak amplitude of 325mV, transmitted power of − 13.05dBm and a noise figure of 9.2dB.
机译:越来越需要开发用于医疗保健监控应用的低功率医疗设备。微电子学,微机电系统(MEMS),电信和微生物医学传感器的最新发展使这种设备的开发和设计成为可能。已经开发了不同的解决方案,例如无线传感器网络(WSN),智能栖息地系统等。在本文中,我们将重点放在言语行为概念上。目的是设计射频发射机架构,以确保实时数据传输。所提出的体系结构基于IEEE 802.15.4a标准,并且在低频段的强制通道3中运行。预期发送器将以0.850 Kbps的数据速率工作并使用BPSK调制。模拟的TX的峰值幅度为325mV,发射功率为− 13.05dBm,噪声系数为9.2dB。