首页> 外文会议>2010 IEEE Students' Technology Symposium (TechSym) >1 V CMOS instrumentation amplifier with high DC electrode offset cancellation for ECG acquisition systems
【24h】

1 V CMOS instrumentation amplifier with high DC electrode offset cancellation for ECG acquisition systems

机译:具有高DC电极偏移消除功能的1 V CMOS仪表放大器,用于ECG采集系统

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

A new circuit topology for a low voltage, low noise instrumentation amplifier (IA) applicable for electrocardiogram signal acquisition system is designed. This circuit is based on the current feedback topology, which is implemented using folded cascode structure at both input and output stages. This helps to sense very low common mode voltage. Achieved input common mode range (ICMR) is 71.43 – 700 mV. DC electrode offset cancellation circuit is realized to remove DC electrode offset along with the low frequency noise. The IA can withstand a high DC electrode offset of ±18 mV. It also has a wide input dynamic range of ±1 µV to ±1 mV. The voltage gain is around 38 dB. It consumes 110 µW power with a supply voltage of 1 V. A high common mode rejection ratio (CMRR) of 80 dB is achieved. The integrated input referred noise is 1.64 µVrms (0.1 Hz – 150 Hz). This design is implemented in 0.18 µm standard CMOS process.
机译:设计了适用于心电图信号采集系统的低电压,低噪声仪表放大器(IA)的新电路拓扑。该电路基于电流反馈拓扑,该电流反馈拓扑在输入和输出级均使用折叠式共源共栅结构实现。这有助于检测非常低的共模电压。达到的输入共模范围(ICMR)为71.43 – 700 mV。直流电极偏移消除电路被实现以去除直流电极偏移以及低频噪声。 IA可以承受±18 mV的高直流电极偏移。它还具有±1 µV至±1 mV的宽输入动态范围。电压增益约为38 dB。它在1 V的电源电压下消耗110 µW的功率。可实现80 dB的高共模抑制比(CMRR)。集成的输入参考噪声为1.64 µV rms (0.1 Hz – 150 Hz)。此设计以0.18 µm标准CMOS工艺实现。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号