MRI image reconstruction under linear and hyperbolic gradients

机译:线性和双曲梯度下的MRI图像重建

获取原文
获取原文并翻译 | 示例

摘要

Nonlinear gradient imaging has advantages of lower peripheral nerve stimulation (PNS), fast imaging rate, and high edge resolution, but the blurring exists in the center part. To resolve this problem, linear gradient x, y is employed together with nonlinear gradient, such as 4-Dimensional Radial In/Out (4DRIO), O-Space, etc. In this paper, we propose a hyperbola method, using x, y gradient and hyperbolic nonlinear gradient x2 − y2. Numerical simulation reveals that hyperbola sampling scheme and iterative algorithm leads to high efficiency in image reconstruction, and hyperbola method shows better reconstruction compare with O-Space technology.
机译:非线性梯度成像具有较低的周围神经刺激(PNS),成像速度快和边缘分辨率高的优点,但是在中心部分存在模糊。为了解决这个问题,将线性梯度x,y与非线性梯度(例如4维径向输入/输出(4DRIO),O-Space等)一起使用。在本文中,我们提出了一种使用x,y的双曲线方法梯度和双曲非线性梯度x 2 -y 2 。数值模拟表明,双曲线采样方案和迭代算法在图像重建中具有很高的效率,与O-Space技术相比,双曲线方法具有更好的重建效果。