【24h】

CFD Approximation of Hemolysis Index

机译:溶血指数的CFD近似值

获取原文
获取原文并翻译 | 示例

摘要

The aim of this paper is to provide a framework for the numerical computation of in vitro hemolysis index over medical devices.The method is based upon the solution of the incompressible Navier- Stokes equations and the algebraic Giersiepen-Wurzinger blood damage equation.A novel interpretation of this equation replaces the computation of blood damage along streamline by a volume integration of a damage function defined in the computational domain.This procedure is validated over 3D test cases and confirms its ability for the classification of medical devices in terms of their innocuousness to blood.
机译:本文旨在为医疗器械体外溶血指数的数值计算提供一个框架。该方法基于不可压缩的Navier-Stokes方程和代数Giersiepen-Wurzinger血液损伤方程的求解。该方程的方程式通过在计算域中定义的损伤函数的体积积分代替了沿流线的血液损伤计算。此过程在3D测试用例上得到了验证,并证实了其对医疗器械的无毒性分类能力。