Comparison of stethoscope bell and diaphragm, and of stethoscope tube length, for clinical blood pressure measurement.

机译:用于临床血压测量的听诊器铃和隔膜以及听诊器管长度的比较。

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

OBJECTIVEududThis study investigated the effect of stethoscope side and tube length on auscultatory blood pressure (BP) measurement.ududMETHODSududThirty-two healthy participants were studied. For each participant, four measurements with different combinations of stethoscope characteristics (bell or diaphragm side, standard or short tube length) were each recorded at two repeat sessions, and eight Korotkoff sound recordings were played twice on separate days to one experienced listener to determine the systolic and diastolic BPs (SBP and DBP). Analysis of variance was carried out to study the measurement repeatability between the two repeat sessions and between the two BP determinations on separate days, as well as the effects of stethoscope side and tube length.ududRESULTSududThere was no significant paired difference between the repeat sessions and between the repeat determinations for both SBP and DBP (all P-values>0.10, except the repeat session for SBP using short tube and diaphragm). The key result was that there was a small but significantly higher DBP on using the bell in comparison with the diaphragm (0.66 mmHg, P=0.007), and a significantly higher SBP on using the short tube in comparison with the standard length (0.77 mmHg, P=0.008).ududCONCLUSIONududThis study shows that stethoscope characteristics have only a small, although statistically significant, influence on clinical BP measurement. Although this helps understand the measurement technique and resolves questions in the published literature, the influence is not clinically significant.
机译:目的 ud ud这项研究调查了听诊器侧面和管长对听诊血压(BP)测量的影响。 ud udMETHODS ud ud研究了32名健康参与者。对于每位参与者,在两次重复练习中分别记录了四次具有不同听诊器特征组合的测量结果(钟形或隔膜侧,标准或短管长度),并在不同的日子分别对一位经验丰富的听众播放了八次Korotkoff录音两次收缩压和舒张压(SBP和DBP)。进行方差分析以研究两次重复会话之间以及两次BP测定在不同天之间的测量可重复性,以及听诊器侧面和管长的影响。 ud udRESULTS ud ud没有显着的配对重复会话之间以及SBP和DBP的重复确定之间的差异(所有P值均大于0.10,但使用短管和隔膜的SBP重复会话除外)。关键结果是,与膜片相比,使用钟罩时的DBP较小但显着较高(0.66 mmHg,P = 0.007),与标准管相比,使用短管时的SBP显着较高(0.77 mmHg ,P = 0.008)。 ud ud结论 ud ud这项研究表明,听诊器的特征对临床血压测量的影响虽然很小,但在统计上却很显着。尽管这有助于了解测量技术并解决了已发表文献中的问题,但这种影响在临床上并不重要。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号