首页> 美国政府科技报告 >Performance evaluation of the solar backscatter ultraviolet radiometer, model 2 (SBUV/2) inflight calibration system
【24h】

Performance evaluation of the solar backscatter ultraviolet radiometer, model 2 (SBUV/2) inflight calibration system

机译:太阳能背向散射紫外辐射计,型号2(sBUV / 2)飞行校准系统的性能评估

获取原文

摘要

The Solar Backscatter Ultraviolet Radiometer, Model 2 (SBUV/2) instruments, as part of their regular operation, deploy ground aluminum reflective diffusers to deflect solar irradiance into the instrument's field-of-view. Previous SBUV instrument diffusers have shown a tendency to degrade in their reflective efficiencies. This degradation will add a trend to the ozone measurements if left uncorrected. An extensive in-flight calibration system was designed into the SBUV/2 instruments to effectively measure the degradation of the solar diffuser (Ball Aerospace Systems Division 1981). Soon after launch, the NOAA-9 SBUV/2 calibration system was unable to track the diffuser's reflectivity changes due, in part, to design flows (Frederick et al. 1986). Subsequently, the NOAA-11 SBUV/2 calibration system was redesigned and an analysis of the first 2 years of data (Weiss et al. 1991) indicated the NOAA-11 SBUV/2 onboard calibration system's performance to be exceeding preflight expectations. This paper will describe the analysis of the first three years NOAA-11 SBUV/2 calibration system data.