首页> 美国政府科技报告 >The effect of truncation on very small cardiac SPECT camerasystems
【24h】

The effect of truncation on very small cardiac SPECT camerasystems

机译:截断对非常小的心脏spECT相机系统的影响

获取原文

摘要

Background: The limited transaxial field-of-view (FOV) of avery small cardiac SPECT camera system causes view-dependent truncationof the projection of structures exterior to, but near the heart. Basictomographic principles suggest that the reconstruction of non-attenuatedtruncated data gives a distortion-free image in the interior of thetruncated region, but the DC term of the Fourier spectrum of thereconstructed image is incorrect, meaning that the intensity scale of thereconstruction is inaccurate. The purpose of this study was tocharacterize the reconstructed image artifacts from truncated data, andto quantify their effects on the measurement of tracer uptake in themyocardial. Particular attention was given to instances where the heartwall is close to hot structures (structures of high activity uptake).Methods: The MCAT phantom was used to simulate a 2D slice of the heartregion. Truncated and non-truncated projections were formed both with andwithout attenuation. The reconstructions were analyzed for artifacts inthe myocardium caused by truncation, and for the effect that attenuationhas relative to increasing those artifacts. Results: The inaccuracy dueto truncation is primarily caused by an incorrect DC component. Forvisualizing theleft ventricular wall, this error is not worse than theeffect of attenuation. The addition of a small hot bowel-like structurenear the left ventricle causes few changes in counts on the wall. Largerartifacts due to the truncation are located at the boundary of thetruncation and can be eliminated by sinogram interpolation. Finally,algebraic reconstruction methods are shown to give better reconstructionresults than an analytical filtered back-projection reconstructionalgorithm. Conclusion: Small inaccuracies in reconstructed images fromsmall FOV camera systems should have little effect on clinicalinterpretation. However, changes in the degree of inaccuracy in countsfrom slice toslice are due to changes in the truncated structures. Thesecan result in a visual 3-dimensional distortion. As with conventionallarge FOV systems attenuation effects have a much more significant effecton image accuracy.

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号