首页> 美国政府科技报告 >Development of Techniques for Detection of Low Molecular Weight Contaminants in Product Water from Water Purification or Water Re-Use Systems.
【24h】

Development of Techniques for Detection of Low Molecular Weight Contaminants in Product Water from Water Purification or Water Re-Use Systems.

机译:水净化或水再利用系统检测产品水中低分子量污染物技术的发展。

获取原文

摘要

A permanganate colorimeter instrument to monitor the quality of MUST product water has been developed and successfully tested at the breadboard level. During the program it was demonstrated that (1) aldehydes are the last of the constituents of MUST RO permeate to be ozonated, (2) residual MUST aldehydes (principally acetaldehyde) rapidly reduce permanganate solutions, and (3) the degree of permanganate reduction may be easily quantitated colorimetrically and is directly proportional to residual MUST water contamination. The colorimeter and required hardware is judged low-cost, rugged, simple, and reliable. (Author)

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号