首页> 美国政府科技报告 >Test and Evaluation of an Energetics Science Incorporated Model 7660 Electrochemical Hydrazines Analyzer
【24h】

Test and Evaluation of an Energetics Science Incorporated Model 7660 Electrochemical Hydrazines Analyzer

机译:Energetics science Incorporated 7660型电化学肼分析仪的测试和评估

获取原文

摘要

An Energetics Science Incorporated Model 7660 MMH analyzer was tested and evaluated for continuous real-time measurement of Hz, MMH, and UDMH in air. Sensitivity to the hydrazines was approximately 0.06 ppm Hz, 0.04 ppm MMH, and 0.025 ppm UDMH. Accuracy and precision, determined at 0.1-ppm concentrations on an analyzer calibrated daily, were between 20% and 40% for the three hydrazines. Analyzer response time increased with age (use) of the sensor. At 100 hours of analyzer operation, the time for 90% of final response was 20 min for Hz; 8 min, MMH; and 1 min, UDMH. Interference studies showed the analyzer responding strongly to ammonia and hydrogen sulfide and less strongly to alcohols and amines. There was also interference from simulated spills of hydraulic fluid, JP-4 fuel, and ECS fluid, with relative analyzer response as high as 6 ppm MMH in the spill area. Interference and response-time problems prevent the analyzer, with the presently used liquid sensor, from being satisfactory for monitoring at or below the threshold limit values for any of the three hydrazines. The instrument is satisfactory for use as a low-ppm leak detector if precautions are taken for distinguishing between the hydrazines and interferents. (Author)

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号