Trueness of a device for intraoral scanning to capture the angle and distance between implants in edentulous mandibular arches

机译:用于口内扫描以捕获无牙颌下颌弓植入物之间的角度和距离的设备的真实性

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

Statement of problem. The digitalization of completely edentulous arches presents limitations because of the lack of anatomic structures for best -fit alignment during the image generation process of the scanner's software program.Purpose. The purpose of this clinical study was to evaluate a new device for intraoral scanning and to analyze its usefulness in capturing the 3-dimensional (3D) position of implants in edentulous mandibular arches.Material and methods. The 3D positions of 40 implants in 10 participants with fixed interim mandibular prostheses were evaluated by comparing 3 scanning techniques. Images were generated in 2 experimental groups, with digital scan bodies (group SC) and with the scanning device (group SD) and in a control group (group CT) in which images were obtained by laboratory scanning of casts produced from splinted impression copings. The standard tessellation language (STL) files were superimposed by using a reverse engineering software program to measure the 3D coordinate system. Variations in implant linear displacements (Ax, p ffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffi Ay, and Az), total 3D displacement ( Dx2+Dy2+Dz2), and angle projections (AqXY, AqXZ, and AqYZ) were statistically analyzed (a=.05). The distances between the implants were also measured. The Spearman correlation coefficient (a=.05) was used to find the correlation between the 3D coordinates and the distances between the implants.Results. Group SD had lower values for linear displacement than group SC; however, this difference was not statistically significant except for implant #2. The overall evaluation showed a significant difference with better accuracy for group SD. Concerning angular displacements, group SD showed lesser angular variation for the 3 projection planes. For the distances between the implants, significant differences were only observed for the "all" assessment in which group SD behaved similarly to group CT, while group SC showed the highest values (P<.05). No correlation was detected between the axes (x, y, and z) and the distances between the implants.Conclusions. The evaluated scanning device led to improved trueness for linear, angular displacements, and distances between implants in mandibular edentulous arches.
机译:问题陈述。完全无牙颌弓的数字化存在局限性,因为在扫描仪软件程序的图像生成过程中缺乏最佳拟合对齐的解剖结构。目的。这项临床研究的目的是评估一种用于口内扫描的新设备,并分析其在捕获无牙颌下颌弓中植入物的 3 维 (3D) 位置方面的有用性。材料和方法。通过比较 3 种扫描技术评估了 10 名使用固定临时下颌假体的参与者的 40 个植入物的 3D 位置。在 2 个实验组中生成图像,使用数字扫描体(SC 组)和扫描设备(SD 组)和对照组(CT 组),其中图像通过实验室扫描由夹板印模产生的石膏获得。标准曲面细分语言(STL)文件通过使用逆向工程软件程序来测量三维坐标系来叠加。对种植体线性位移(Ax、p ffiffiff还测量了植入物之间的距离。使用Spearman相关系数(a=.05)求出3D坐标与植入物之间距离之间的相关性。结果。SD组的线性位移值低于SC组;然而,除了植入物 #2 外,这种差异没有统计学意义。总体评价显示,SD组的角位移差异显著,精度更高。 关于角位移,SD组在3个投影平面上表现出较小的角变化。对于植入物之间的距离,仅在“所有”评估中观察到显着差异,其中 SD 组的行为与 CT 组相似,而 SC 组显示最高值 (P<.05)。未检测到轴(x、y 和 z)与植入物之间的距离之间的相关性。结论。评估的扫描设备提高了下颌无牙颌弓中线性、角度位移和植入物之间距离的真实度。

著录项

相似文献

  • 外文文献
  • 中文文献
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号