首页> 外文期刊>Medical Physics >Validation of 3D reconstructions of a mimicked femoral artery with an ultrasound imaging robotic system
【24h】

Validation of 3D reconstructions of a mimicked femoral artery with an ultrasound imaging robotic system

机译:用超声成像机器人系统验证模拟股动脉的3D重建

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Purpose: The degree of stenosis is the most important criterion to assess peripheral arterial disease manifested by atherosclerosis mainly in lower limb arteries. Ultrasound (U.S.) imaging offers low-cost, safe, and convenient options to evaluate this disease, but most U.S. freehand approaches cannot optimally locate stenoses and map lower limb arterial geometries. A 3D-U.S. imaging robotic system that can control and standardize image acquisition by scanning typically encountered diseased arterial lower limb segments is presented and validated with phantoms. Methods: A Z-phantom calibration procedure was used to characterize spatial transformation of the U.S. probe image plane for different clinical image acquisition settings. Moreover, the accuracy of the calibration transform to reconstruct a lower-limb-mimicking vessel geometry was evaluated with a vascular phantom. Results: A 3D calibration precision of 0.47 +- 0.27 mm was achieved. Reconstruction errors were less than 1.74 + 0.08 mm in all 3D vessel representations and the cross-sectional areas of each image section were close to those of gold standard phantom measures. The best reconstruction accuracy (smallest error) was 0.40 +- 0.03 mm. Conclusion: Altogether, these results demonstrate the potential of the robotic scanner to adequately represent lower limb vessels for the clinical evaluation of stenoses.
机译:目的:狭窄程度是评估主要在下肢动脉中以动脉粥样硬化为表现的外周动脉疾病的最重要标准。超声(U.S.)成像提供了低成本,安全和方便的方法来评估这种疾病,但是大多数美国徒手绘制方法无法最佳地定位狭窄并绘制下肢动脉的几何形状。 3D-U.S。提出了一种成像机器人系统,该系统可以通过扫描通常遇到的患病的动脉下肢节段来控制和标准化图像采集,并通过幻像进行了验证。方法:使用Z幻像校准程序来表征针对不同临床图像采集设置的美国探头图像平面的空间变换。此外,用血管模型评估了校准变换重建下肢模拟血管几何形状的准确性。结果:实现了0.47±0.27 mm的3D校准精度。在所有3D血管图示中,重建误差均小于1.74 + 0.08毫米,并且每个图像截面的横截面面积均接近金标准幻象测量的横截面面积。最佳的重建精度(最小的误差)为0.40±0.03 mm。结论:总而言之,这些结果表明机器人扫描仪有潜力充分代表下肢血管,用于狭窄的临床评估。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号