COMPARISON OF TWO METHODS FOR EVALUATION OF THE IMAGE QUALITY OF LUMBAR SPINE RADIOGRAPHS

机译:

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Two methods for visual evaluation of image quality of clinical radiographs have been compared. In visual grading analysis (VGA) specified anatomical structures in an image are visually compared with the same structures in a reference image, and in a free-response forced error (FFE) experiment—an extension of conventional ROC (receiver operating characteristics) analysis—the objective is to localise known lesions correctly. The spatial resolution and noise of digitised clinical radiographs of the lumbar spine were altered by image processing, and pathological structures were added to the images for the FFE experiment. The images were printed to film and evaluated by seven European expert radiologists using VGA and FFE. The results of these two different methods showed a very good agreement. VGA methodology can be made as solid as the FFE experiment for evaluating image quality. The simplicity of VGA makes it very suitable for implementation in clinical practice.

著录项

  • 来源
    《Radiation protection dosimetry》 |2000年第2期|165-168|共4页
  • 作者单位

    Department of Radiation Physics at Malmo, Lund University Malmo University Hospital, SE-205 02 Malmo, Sweden;

    GSF-National Research Center for Environment and Health, D-85764 Neuherberg, Germany;

    Department of Radiation Physics, Goteborg University Sahlgrenska University Hospital, SE-413 45 Goteborg, Sweden;

  • 收录信息 美国《科学引文索引》(SCI);美国《化学文摘》(CA);
  • 原文格式 PDF
  • 正文语种 英语
  • 中图分类 TL72;
  • 关键词

获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号