Dose reduction during CT coronary angiography

机译:CT冠状动脉造影期间的剂量减少

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

Dose reduction during CT coronary angiography Dose delivery during CT coronary angiography with retrospective ECG gating is high especially due to the important slice overlapping. Optimization of the acquisition parameters is necessary to reduce patient exposure. First, the height of the scan field should be limited to the heart. Both kV and mA should be adjusted based on patient morphology. ECG gated exposure modulation with mA reduction during systole, a technique most applicable for patients with slow and regular heart rate, can result in a dose reduction up to 50. The use of prospective ECG gating can also reduce patient dose. This technique also requires patients with slow and regular heart rate.
机译:CT 冠状动脉造影期间的剂量减少 CT 冠状动脉造影和回顾性心电图门控期间的剂量递送很高,特别是由于重要的切片重叠。为了减少患者暴露,必须优化采集参数。首先,扫描区域的高度应限制在心脏。kV 和 mA 都应根据患者形态进行调整。心电图门控暴露调节在收缩期降低 mA,一种最适用于心率缓慢且规律的患者的技术,可导致剂量减少高达 50%。使用前瞻性心电图门控也可以减少患者剂量。该技术还要求患者心率缓慢且规律。