Assessing the validity of oscillometric device for?blood pressure measurement in a large population-based epidemiologic study

机译:在一项基于人群的大型流行病学研究中评估示波器用于血压测量的有效性

获取原文
获取原文并翻译 | 示例

摘要

Abstract To evaluate the accuracy of the Omron HBP-1300, an automatic device for blood pressure (BP) measurement, compared with the standard auscultatory method with a mercury sphygmomanometer, we conducted a large population-based survey. From 2012 to 2015, a total of 500,000 participants were randomly selected from the 262 urban cities and rural counties in China, of them BP was measured three times simultaneously with mercury sphygmomanometer and Omron HBP-1300 in 37,816 participants. Bias was estimated as the average of the auscultatory measurements minus the average of the oscillometric measurements, with its corresponding 95 limits of agreement. The Association for the Advancement of Medical Instrumentation standards and British Hypertension Society protocols were used to evaluate the accuracy of the oscillometric device. The results showed that the Omron underestimated systolic BP by 1.0爉m Hg (95 limits of agreement: 9.2, 11.2) and underestimated diastolic BP by 0.2爉m Hg (95 limits of agreement: 8.5, 8.1). The bias in systolic BP values was related to heart rate, systolic BP, and pulse pressure level, while the bias in diastolic BP was related to the heart rate and pulse pressure level. Omron HBP-1300 passed the Association for the Advancement of Medical Instrumentation standards and British Hypertension Society protocol requirements. The sensitivity and specificity of the Omron HBP-1300 to detect hypertension were 86.2 and 98.0, respectively. Our results showed that the Omron HBP-1300 could be used for measuring BP in large epidemiology studies. The bias of BP measurement was related to the age, systolic BP, and pulse pressure level. Highlights " Omron HBP-1300 underestimated systolic/diastolic blood pressure by 1.0/0.2爉m Hg. " Omron passed the Association for the Advancement of Medical Instrumentation standards and British Hypertension Society protocol requirements. " The sensitivity and specificity of the Omron HBP-1300 to detect hypertension were 86.2 and 98.0, respectively. " Omron HBP-1300 could be used for measuring blood pressure in large epidemiology studies.
机译:摘要 为了评估欧姆龙HBP-1300(一种用于血压(BP)的自动测量设备,与使用水银血压计的标准听诊方法相比,我们进行了大规模的人群调查。从2012年到2015年,从全国262个城市和农村县随机抽取了50万名受试者,其中37816名受试者使用水银血压计和欧姆龙HBP-1300同时测量了3次血压。偏倚估计为听诊测量值的平均值减去示波测量值的平均值,其相应的 95% 一致性限。医疗仪器促进协会标准和英国高血压学会协议用于评估示波器的准确性。结果显示,欧姆龙低估了收缩压1.0爉m汞柱(95%一致限:9.2,11.2),低估了舒张压0.2爉m汞柱(95%一致限:8.5,8.1)。收缩压值的偏倚与心率、收缩压和脉压水平有关,而舒张压的偏倚与心率和脉压水平有关。欧姆龙HBP-1300通过了医疗器械促进会标准和英国高血压协会协议要求。欧姆龙 HBP-1300 检测高血压的敏感性和特异性分别为 86.2% 和 98.0%。我们的结果表明,欧姆龙HBP-1300可用于大型流行病学研究中的血压测量。血压测量的偏倚与年龄、收缩压和脉压水平有关。亮点 “ 欧姆龙HBP-1300低估了收缩压/舒张压1.0/0.2爉m汞柱。欧姆龙 HBP-1300 检测高血压的敏感性和特异性分别为 86.2% 和 98.0%。“ 欧姆龙HBP-1300可用于大型流行病学研究中的血压测量。

著录项

相似文献

  • 外文文献
  • 中文文献
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号