Noninvasive Quantitation of Liver Iron in Dogs with Hemochromatosis Using Dual‐Energy CT Scanning

机译:使用双连字符能量 CT 扫描无创定量血色素沉着症犬的肝铁

获取原文
获取外文期刊封面目录资料

摘要

The concentration of iron deposited.in the livers of two dogs with experimentally induced iron overload was determined by use of dual energy computerized tomographic (CT) scanning. A phantom was constructed, containing known amounts of iron-dextran solutions. CT scans of the phantoms, at 80 and 120 kVp, corrected for the response of water, showed a linear relationship between known iron concentrations and difference in CT number at the two scanning energies, with a change of 24 H units per 1000 mg iron. Using the graph of this linear relationship, the amount of iron in dog liver was predicted, compared with the amount of iron measured from biopsy specimens, and analyzed by neutron activation analysis. A close correlation existed between predicted liver iron and measured iron concentration (r= 0.99). Dual-energy CT scanning appears to provide an accurate, noninvasive method of quantitating liver iron.
机译:通过使用双能计算机断层扫描 (CT) 测定两只实验诱导铁过载的狗肝脏 deposited.in 铁的浓度。构建了一个模型,其中包含已知量的右旋糖酐铁溶液。在80和120 kVp下对模型进行CT扫描,校正了水的响应,显示已知的铁浓度与两种扫描能量下的CT值差异之间存在线性关系,每1000 mg%铁的变化为24 H单位。利用这种线性关系图,预测狗肝脏中的铁含量,与活检标本测得的铁含量进行比较,并通过中子活化分析进行分析。预测的肝铁与测得的铁浓度之间存在密切相关性(r=0.99)。双能 CT 扫描似乎提供了一种准确、无创的肝铁定量方法。